Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 29-2: CAL ADJ /5100 INSTRUMENT: Fluke 79-2: CAL ADJ /5100 DATE: 06-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 2.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 60 CONFIGURATION: Fluke 5100B ============================================================================= # # Source: # Fluke 79/29/ Series II Service Manual # (PN 896209 July 1991, Rev. 2, 1/94, Errata No. 1 11/94) # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5100B accuracy file contains 6 month specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 6 month specifications of the 5100B are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- W 1.002 HEAD EQUIPMENT SETUP 1.003 DISP Disconnect the test leads if attached. Remove the four 1.003 DISP phillips-head screws from the bottom of the case. Turn 1.003 DISP the instrument face-up, and grasp the top cover at both 1.003 DISP sides of the input connectors. Then, pull the top cover 1.003 DISP from the UUT. 1.004 DISP Turn the UUT on. 1.004 DISP Ambient temperature: 18C - 28C. 1.004 DISP Relative humidity: less than 80%. 1.004 DISP Allow the UUT to stabilize. 1.005 HEAD DC VOLTAGE CALIBRATION 1.006 DISP Select the V DC function on the UUT. 1.007 DISP Connect the 5100B and UUT as follows: 1.007 DISP [32] 5100B OUTPUT HI to UUT V Ohm DIODE 1.007 DISP [32] 5100B OUTPUT LO to UUT COM 1.008 5100 3.5V S 2W 1.009 DISP Adjust R21 for a UUT reading of 3.500+/-0.001V. 1.010 5100 * S 1.011 HEAD AC VOLTAGE CALIBRATION 1.012 DISP Select the V AC function on the UUT. 1.013 5100 3.5V 100H S 2W 1.014 DISP Adjust R34 for a UUT reading of 3.500+/-0.002V. 1.015 END